Invacare 1900 Instrukcja Użytkownika Strona 8

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 7
SECTION 1—GENERAL GUIDELINES
1900/1900S Bath Lift 8 Part No. 1148113
Performing Maintenance
RefertoMaintenanceandTroubleshootingonpage 17foramaintenancescheduleand
procedures.
Regularmaintenanceofliftsandaccessoriesisnecessarytoassureproperoperation.
Castersandaxleboltsrequireinspectionseverysixmonthstocheckfortightness,wear,
debris(suchashairanddirt)andthattheyrollfree.
Afterthefirsttwelvemonthsofoperation,inspectallglides,rollers,andfastenersfor
wear.Ifthemetalisworn,thepartsMUSTbereplaced.Makethisinspectioneverysix
monthsthereafter.
Theelectricmotorissealedatthefactoryandifserviceisrequired, themotorunitMUST
bereturnedtothefactoryforrepair.DONOTattempttoopenthemotororobtainlocal
serviceasthiswillVOIDthewarrantyandmayresultindamageandacostlyrepair.
ConsultyourdealerorInvacareforfurtherinformation.
Electrical - Grounding Instructions
DONOT,underanycircumstances,cutorremovetheroundgroundingprongfromany
plug.Somedevicesareequippedwiththreeprong(grounding)plugsforprotection
againstpossibleshockhazards.Whereatwoprongwallreceptacleisencountered,itis
thepersonalresponsibilityandobligationofthecustomertocontactaqualified
electricianandhavethetwoprongreceptaclereplacedwithaproperlygroundedthree
prongwallreceptacleinaccordancewiththeNationalElectricalCode.Ifyoumustusean
extensioncord,useonlyathreewireextensioncordhavingthesameorhigherelectrical
ratingasthedevicebeingconnected.Inaddition,InvacarehasplacedRED/ORANGE
WARNINGTAGSonsomeequipment.DONOTremovethesetags.Carefullyread
battery/batterychargerinformationpriortoinstalling,servicingoroperatingyourlift.
Pinch Points
WARNING
Pinch points exist at base of lift. Be careful, injury could occur.
FIGURE 1.1 Pinch Points
Pinch
Points
Lift
Base
Caster
Caster
Casters